Gas Chromatography–Mass Spectrometry (GC-MS)

Instrument detail:

ModelAgilent Technologies 7890 B GC System / 5977 A MSD (Mass spectral Detector) / Electron Ionization method / NIST Library for identification of compounds.
MakeAgilent Technologies 7890B GC system/5977A MSD
SoftwareMass hunter
Column TypeCapillary Column HP-5 (Crossbond, 5% Phenyl and  95% dimethylpolysiloxane)
Column length30m
Column id250µm
Total GC Run Time40.50 minutes

MS Conditions:

Mass Range20-550amu
Mode Full Scan Mode
Type of IonizationElectron Ionization (EI)
Electron energy 70ev
Transfer line and source temperature 230ºC
LibraryNIST 2014

Applications:

This facility can be used for the identification of semi-volatile and volatile organic compounds (minimum sample size required is 2 µl).

Charges (inclusive of GST)

TestAmount
GCMS Analysis and preparation of reportRs. 3,540/- (per sample)

 